3 loading the appliance – Memmert ICP Compressor-cooled incubators User Manual
Page 21

D30374 | Date 10/2014
21
Operation and control
5.3 Loading the appliance
Warning!
When loading the appliance with an unsuitable load, poisonous
or explosive vapours or gases may be produced. This could cause
the appliance to explode, and persons could be severely injured or
poisoned. The appliance may only be loaded with materials which do
not form any toxic or explosive vapours when heated up and cannot
ignite (see also Intended use on page 8). If there is any doubt as
to the composition of materials, they must not be loaded into the
appliance.
Caution:
Check the chamber load for chemical compatibility with the materials of the appliance
(see page 10).
Insert the sliding steel grids or sliding shelves. The maximum number or grids / shelves and the
load capacity are specified in the technical data overview from page 12 .
The chamber must not be
loaded too tightly, so that
proper air circulation in the
interior is guaranteed. Do
not place any of the chamber
load on the floor, touching
the side walls or right below
the ceiling of the chamber
(Fig. 12, see also the "correct
loading" sticker on the appli-
ance).
In case of improper loading
(chamber loaded too tightly),
the set temperature may
be exceeded or it may take
longer until it is reached.
To achieve a suitable heat output, the type of drawer used – grid or shelf – has to
be set in the menu under
SETUP
(see page 39).
Loading – Models 30 - 1060
Open / close the door
E24899
Memmert GmbH + Co. KG | P.O. Box 1720 | D-91107 Schwabach | Tel. +49 (0) 9122 / 925 - 0 | Fax +49 (0) 9122 / 145 85 | E-Mail: [email protected]
www.memmert.com | www.atmosafe.net
Fig. 12
Correct placement of the chamber load