Page 27: Alarm, Display shows when the programme should end, Restart the programme as described, Electronic temperature monitoring (tww), Automatic temperature monitor (asf)

background image

D30374 | Date 10/2014

27

Operation and control

3. Press the confirmation key to

confirm. The programme is

cancelled.

Hauptschalter

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung

getroffen hat.

Die hier gezeigt Variante, ist die von uns empfohlene Richtung.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept

13:44

End

Test 012

12.11.2012

A cancelled programme cannot be resumed at the point it was cancelled. It must be

restarted from the beginning.

End of programme

End of programme

The

End

display shows when the

programme should end.

13:44

End

Test 012

12.11.2012

restart the programme as described

select another programme for processing in menu mode (see page 44) and run it as

described.

return to manual mode. To do

so, reactivate it by pressing the

activation key next to the status

display, then turn the turn con-

trol until the hand symbol

is

is

is

highlighted in colour and press

the confirmation key.

5.5 Monitoring function

5.5.1 Temperature monitoring

The appliance is equipped with a multiple overtemperature protection in accordance with

DIN 12 880. This serves to avoid damage to the chamber load and/or appliance in case of a

malfunction:

electronic temperature monitoring (TWW)

automatic temperature monitor (ASF)

mechanical temperature limiter (TB)

The monitoring temperature of the electronic

temperature monitoring is measured via a separate Pt100 tem-

perature sensor in the interior. Temperature monitoring settings

are made via the

ALARM

display. The settings made apply to all

operating modes.

Manual mode

12.09.2012

13:44

ALARM

max

38

.5°C

2.0

K

min

36

.5°C

auto

+-