beautypg.com

6 setting up, 1 prerequisites – Memmert ICP Compressor-cooled incubators User Manual

Page 17

background image

D30374 | Date 10/2014

17

Delivery, transport and setting up

3.6 Setting up

Warning!

Due to its centre of gravity, the appliance can fall over to the front

and injure you or other people. Always attach the appliance to a wall

with the tilt protection (see page 18). If this is not possible due to

spatial constrictions, do not put the appliance into operation and do

not open the door. Please contact Memmert customer service (see

page 2).

The appliance must only be set up on the fl oor.

3.6.1 Prerequisites

The installation site must be flat and horizontal and must be able to reliably bear the weight

of the appliance (see Technical data on page 12). Do not place the appliance on a flamma-

ble surface.
Depending on the model (see nameplate), a 230 V or 115 V power connection must be avail-

able at the installation site.
The distance between the wall and the rear of the appliance must be at least 15 cm. The clear-

ance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby

appliances must not be less than 5 cm (Fig. 8). Sufficient air circulation in the vicinity of the

appliance must be guaranteed at all times. Do not place any objects in front of the fan open-

ing of the cooling unit at the front of the appliance.
For appliances with castors, always direct them to the front.

≥ 5 cm

≥ 15 cm

≥ 20 cm

≥ 5 cm

Fig. 8

Minimum clearance from walls and ceiling