Malfunctions, warning and error messages

Description

Cause

Action

See

Appliance does not heat

up

The mechanical

temperature lim-

iter (TB) perma-

nently switched

off heating.

1. Leave the appliance to cool

down.

2. Reset the TB. To do so, press

the red button at the rear

side of the appliance until

it locks into place with an

audible click.

If the alarm continues: Contact

customer service

page 2

6.2 Malfunctions, operating problems and appliance errors

Error description

Cause of errors

Rectifying errors

See

Displays are dark

External power supply

was interrupted

Check the

power supply

page 19

Miniature fuse, appliance

fuse or power module

faulty

Contact customer

service

page 2

Displays cannot be

activated

Appliance locked by USER

ID

Unlock with USER ID

Page 45

The appliance is in pro-

gramme, timer or remote

control mode (mode "Write"

or "Write + Alarm")

Wait until the end of

the programme or

timer mode or switch

off the remote control

Displays suddenly look

different

Appliance is in "wrong"

mode

Change to operating

or menu mode by

pressing the

MENU

key

Display T:E-3 in the

temperature display

TEMP

T:E-3

Set 37.0

°C

37

.4

°C

Temperature operating sen-

sor defective. The monitor-

ing sensor takes over the

measurement function.

The appliance can

be temporarily

operated

Contact customer

service as soon as

possible

page 2