2 programme example with door locking – Memmert AtmoCONTROL User Manual
Page 25

D24042 | Date 09/2014
25
AtmoCONTROL
4.4.2 Programme example with door locking
1
2
5
8
6
3
4
Tolerance band
1,7 K/min
7
95 °C
20 °C
45 °C
12:00 h ≥ 0:30 h
1
2
3
5
6
Fig. 9
The door is locked at the beginning of the programme (1). Then, the appliance heats up to
Fig. 9
Fig. 9
95.0 °C (2) and maintains this temperature for 12 hours (3). Subsequently, the temperature is
The door is locked at the beginning of the programme (1). Then, the appliance heats up to
The door is locked at the beginning of the programme (1). Then, the appliance heats up to
lowered (5) for 30 minutes (4) to 45.0
C (2) and maintains this temperature for 12 hours (3). Subsequently, the temperature is
C (2) and maintains this temperature for 12 hours (3). Subsequently, the temperature is
°C and then, the door is opened (6). The setting "
C (2) and maintains this temperature for 12 hours (3). Subsequently, the temperature is
C (2) and maintains this temperature for 12 hours (3). Subsequently, the temperature is
SPWT
on" (8) ensures that the door is opened only when the temperature really has dropped to
C and then, the door is opened (6). The setting "
C and then, the door is opened (6). The setting "
45.0 °C, even if this takes longer than 30 minutes. Below the temperature change is shown in
on" (8) ensures that the door is opened only when the temperature really has dropped to
on" (8) ensures that the door is opened only when the temperature really has dropped to
K/min (7).
C, even if this takes longer than 30 minutes. Below the temperature change is shown in
C, even if this takes longer than 30 minutes. Below the temperature change is shown in