Function, Adjustment options, Atmocontrol – Memmert AtmoCONTROL User Manual
Page 17: Hold humidity, Maintains a specific humidity for a specific time, Name of programme segment, Duration (time or infinite ∞), Humidity value to be main- tained, Tolerance value above/below, Alarm if limits are exceeded

D24042 | Date 09/2014
17
AtmoCONTROL
Depiction
in icon bar
Meaning
Depiction on
editor thread
Function and
adjustment options
?
15
STAND BY
15
Hold
humidity
5
.
0
%rh
Tolerance band
+
-
Alarm
Safe
on off
70
.0
%rh
12
h
:
00
m
Hold
0
.
0
%rh
Function
Maintains a specific humidity for a
specific time.
Adjustment options
Adjustment options
•
Name of programme segment
1
•
Duration (time or infinite ∞)
•
Humidity value to be main-
tained
•
Tolerance value above/below
•
Alarm if limits are exceeded
•
Safe
2
?
15
STAND BY
15
Change
humidity
2
h
:
30
m
40
.7
%rh
70
.6
%rh
5
.
0
%rh
5
.
0
%rh
+
-
SPWT
on off
Tolerance band
Increase
Function
Increases or decreases humidity
over a specific time to a specific
value.
Adjustment options
Adjustment options
•
Name of programme segment
1
•
Duration
•
Target (setpoint) humidity
•
Tolerance value above/below
•
SPWT
3
1
When run, this is displayed in the status bar of the appliance
2
When Safe is "on", it is ensured that the value really is maintained within the tolerance band as long as
specified, and only then is the programme continued (this is sensible for sterilisers, for example). If the
actual values leaves the tolerance band, the clock timer starts again from the beginning.
3
SPWT: Setpoint wait. If this is "on", the programme sequence is not continued before the setpoint
value is reached, even if the set time has already expired. If this is "off", the programme sequence is
continued after the set time has expired, irrespective of whether the setpoint value was reached or not.