Function, Adjustment options, Atmocontrol – Memmert AtmoCONTROL User Manual

Page 17: Hold humidity, Maintains a specific humidity for a specific time, Name of programme segment, Duration (time or infinite ∞), Humidity value to be main- tained, Tolerance value above/below, Alarm if limits are exceeded

background image

D24042 | Date 09/2014

17

AtmoCONTROL

Depiction

in icon bar

Meaning

Depiction on

editor thread

Function and

adjustment options

?

15

STAND BY

15

Hold

humidity

5

.

0

%rh

Tolerance band

+

-

Alarm

Safe

on off

70

.0

%rh

12

h

:

00

m

Hold

0

.

0

%rh

Function

Maintains a specific humidity for a

specific time.

Adjustment options

Adjustment options

Name of programme segment

1

Duration (time or infinite ∞)

Humidity value to be main-

tained

Tolerance value above/below

Alarm if limits are exceeded

Safe

2

?

15

STAND BY

15

Change

humidity

2

h

:

30

m

40

.7

%rh

70

.6

%rh

5

.

0

%rh

5

.

0

%rh

+

-

SPWT

on off

Tolerance band

Increase

Function

Increases or decreases humidity

over a specific time to a specific

value.

Adjustment options

Adjustment options

Name of programme segment

1

Duration

Target (setpoint) humidity

Tolerance value above/below

SPWT

3

1

When run, this is displayed in the status bar of the appliance

2

When Safe is "on", it is ensured that the value really is maintained within the tolerance band as long as

specified, and only then is the programme continued (this is sensible for sterilisers, for example). If the

actual values leaves the tolerance band, the clock timer starts again from the beginning.

3

SPWT: Setpoint wait. If this is "on", the programme sequence is not continued before the setpoint

value is reached, even if the set time has already expired. If this is "off", the programme sequence is

continued after the set time has expired, irrespective of whether the setpoint value was reached or not.