Removing a parameter icon from the editor thread – Memmert AtmoCONTROL User Manual
Page 14

14
D24042 | Date 09/2014
AtmoCONTROL
To create a programme, drag the individual parameter icons onto the editor thread one after
another in the desired order, while holding down the left mouse button (Fig. 3 and Fig. 4). To
assist the correct positioning, a red insertion mark is displayed at the insertion position. With
the zoom icons in the toolbar list (see section 3.2.2 on page 9) or with the mouse wheel,
you can zoom in or out of the display or have the entire programme displayed.
180
.0
°C
10
h
:
00
m
21
.0
°C
180
.0
°C
Ramp 05
11
h
:
30
m
Fig. 3
Drag the parameter icon (in this case a tem-
perature change) onto the editor thread while
holding the mouse button down
40
.7
%rh
70
.6
%rh
2
h
:
30
m
Increase
180
.0
°C
10
h
:
00
m
21
.0
°C
180
.0
°C
Ramp 05
11
h
:
30
m
0
%
LIGHT
Fig. 4
Drag other parameters – light and a humidity
change in this case – onto the editor thread.
A red insertion mark helps you to find the
correct position.
The temperature icons (change/hold temperature) may only be placed on the upper editor
thread, humidity and pressure icons only on the lower one.
The meaning of the individual icons and the adjustment options are described from page
16. You can find some simple programme examples from page 24.
Removing a parameter icon from the editor thread
Removing a parameter icon from the editor thread
To remove a parameter
icon (and therefore its
function) from an edi-
tor thread – if you have
inserted it by mistake, for
example – select it and,
with the mouse button
depressed, move it to the
recycle bin symbol on the
lower right (Fig. 5).
To delete a parameter icon from an editor thread, drag the icon,