Function, Adjustment options, Atmocontrol – Memmert AtmoCONTROL User Manual
Page 18: Hold pres- sure, Maintains a specific pressure for a specific time, Name of programme segment, Duration (time or infinite ∞), Pressure to be maintained, Tolerance value above/below, Alarm if limits are exceeded

18
D24042 | Date 09/2014
AtmoCONTROL
Depiction
in icon bar
Meaning
Depiction on
editor thread
Function and
adjustment options
?
15
STAND BY
15
Hold pres-
sure
+
-
Tolerance band
Alarm
Safe
on off
hold p
20
mb
10
mb
2
h
:
30
m
500
mb
Function
Maintains a specific pressure for a
specific time.
Adjustment options
Adjustment options
•
Name of programme segment
1
•
Duration (time or infinite ∞)
•
Pressure to be maintained
•
Tolerance value above/below
•
Alarm if limits are exceeded
•
Safe
2
?
15
STAND BY
15
Change
pressure
+
-
Tolerance band
1100
mb
decrease
2
h
:
30
m
500
mb
10
mb
20
mb
SPWT
on off
Function
Increases or decreases pressure
over a specific time to a specific
value.
Adjustment options
Adjustment options
•
Name of programme segment
1
•
Duration
•
Target (setpoint) pressure
•
Tolerance value above/below
•
SPWT
3
1
When run, this is displayed in the status bar of the appliance
2
When Safe is "on", it is ensured that the value really is maintained within the tolerance band as long as
specified, and only then is the programme continued (this is sensible for sterilisers, for example). If the
actual values leaves the tolerance band, the clock timer starts again from the beginning.
3
SPWT: Setpoint wait. If this is "on", the programme sequence is not continued before the setpoint
value is reached, even if the set time has already expired. If this is "off", the programme sequence is
continued after the set time has expired, irrespective of whether the setpoint value was reached or not.