4 available parameters – Memmert AtmoCONTROL User Manual
Page 16

16
D24042 | Date 09/2014
AtmoCONTROL
4.1.4 Available parameters
Below, all the parameter icons with their adjustment options are shown.
Which parameters are available to adjust the programme depends on the appliance for
which a programme is to be created. Only those parameters are available that the appli-
ance is able to implement. For appliances without humidity regulation, for example, no
humidity icon is available. The respective adjustment options (temperature ranges etc.) are
appliance-specific.
Broad parameter representation
parameter representation
Depiction
in icon bar
Meaning
Depiction on
editor thread
Function and
adjustment options
?
15
STAND BY
15
Hold tem-
perature
Tolerance band
Function
Maintains a set temperature for a
specific time.
Adjustment options
Adjustment options
•
Name of programme segment
1
•
Duration (time or infinite ∞)
•
Temperature to be maintained
•
Tolerance value above/below
•
Alarm if limits are exceeded
•
Safe
2
(For programme example, see page
24)
?
15
STAND BY
15
Change
tempera-
ture
0,1 K/min
Tolerance band
Function
Increases or decreases temperature
over a specific time to a set value.
Adjustment options
Adjustment options
•
Name of programme segment
1
•
Duration
•
Target (setpoint) temperature
•
Tolerance value above/below
•
SPWT
3
(For programme example, see page
25)
1
When run, this is displayed in the status bar of the appliance
2
When Safe is "on", it is ensured that the value really is maintained within the tolerance band as long as
specified, and only then is the programme continued (this is sensible for sterilisers, for example). If the
actual values leaves the tolerance band, the clock timer starts again from the beginning.
3
SPWT: Setpoint wait. If this is "on", the programme sequence is not continued before the setpoint
value is reached, even if the set time has already expired. If this is "off", the programme sequence is
continued after the set time has expired, irrespective of whether the setpoint value was reached or not.