Calculating the total magnification/field of view diameter
Parameter
MO
Magnification of the additional objective
ME
Magnification of eyepiece
z
Magnification changer position
NFOV Field number of the eyepiece. Field numbers are printed on
the eyepieces: 10Ч/23, 16Ч/16, 20Ч/12, 10Ч/23B, 16×14B,
25×/9.5B, 40×6B.
Example
MO
Additional objective 1.6×
ME
20×/12 eyepiece
z
Zoom position 4.0
Magnification in the binocular tube
MTOT VIS = MO × ME × z
or
1.6 Ч 20 Ч 4 = 128
Calculation example: Field of view diameter in the specimen
∅ OF:
NFOV
=
12
= 1.9 mm
MO × z
1.6 Ч4
See also other documents in the category Leica Biosystems Equipment:
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)