Troubleshooting – Leica Biosystems Bond Oracle HER2 IHC System User Manual
Page 21

Page 21 of 23
English
Leica Biosystems Bond Oracle HER2 IHC System Instructions for Use TA9145 EN-CE-Rev_E 18/06/2013
English
Troubleshooting
Problem
Probable Cause
Remedial Action
No immunohistochemical
staining
Run aborted prior to
completion
Using BOND software, confirm the presence of
any reportable errors during the staining run and
address as instructed by the BOND software.
Incorrect protocol selection
Ensure appropriate default to *IHC Protocol H
in
the staining protocol field of the Add slide dialog.
Inadequate
deparaffinization of slides
Ensure *Dewax mode is selected in the
Preparation field of the Add slide dialog.
Inappropriate bulk reagents
dispensed
Ensure all BOND reagents have been allocated
to appropriate bulk containers and placed into
appropriate positions on the instrument.
Contamination of BOND Wash
Solution with sodium azide
Use fresh BOND Wash Solution prepared to
appropriate working strength.
Weak specific
immunohistochemical
staining
Inappropriate epitope
retrieval
Ensure appropriate BOND Epitope Retrieval reagents
have been allocated into correct bulk containers, and
BOND software has defaulted to the appropriate
epitope retrieval protocol,
*HIER 25 min with *ER1 (97).
Inappropriate fixation
or processing of test
specimen
Ensure a formalin-based fixative is used and
that processing schedules are suitable for the
specimen undergoing testing.
Bond Oracle HER2 IHC System
is being used outside its expiry
date
Ensure the Bond Oracle HER2 IHC System used is
within its specified expiry date.
Excessive specific
immunohistochemical
staining
Inappropriate epitope
retrieval
Ensure appropriate BOND Epitope Retrieval
reagents have been allocated into appropriate bulk
containers, and the BOND software has defaulted
to *HIER 25 min with ER1 (97).
Variation in fixation
Ensure a formalin-based fixative is used and
that processing schedules are suitable for the
specimen undergoing testing. If possible, retest
case using another block. If this is not possible,
assess the areas which show best fixation
patterns in conjunction with a corresponding
H&E stained section.