Safety precautions – Leica Biosystems EG1160 User Manual
Page 8
8
Instruction manual V 4.1 – 05/2001
-
The instrument may be used by qualified personnel only.
-
To avoid any operating mistakes from the beginning, it is necessary to
read the operating instructions to familiarize oneself with the technical
details.
-
Illumination: Before replacing the lamp, the tissue embedding center must
be turned off with the main switch.
-
To clean the condenser fins, switch the instrument off with the main
switch and disconnect it from mains.
-
When operating, the paraffin reservoir, mold warmer, cassette bath, work
area and forceps holder are all hot.
Caution: Risk of burning!
Note: Combustible and flammable substances must not be stored near
the instrument!
-
Paraffin is flammable and should be handled with care. Spillage should
be avoided. Paraffin on the surfaces must not be removed with sharp
tools as it would ruin the coating. It should be avoided to allow xylene to
react on all surfaces.
Caution:
Xylene is a flammable organic solvent. Its flash point is between 27 and
32°C. Xylene vapours are heavier than air and can easily catch fire on
hot surfaces or sparks even over a greater distance.
-
To replace the fuses, the power cord must be disconnected from the
power supply. Only fuses that are accessible from outside can be re-
placed. If you need further information, please contact your local Leica
sales office or dealer.