Technical data – Leica Biosystems EG1160 User Manual
Page 6
![background image](/manuals/565394/6/background.png)
6
Instruction manual V4.0 – 4/98
Power supply:
115 V / 60 Hz
230 V / 50 Hz
Fuses:
115V/60Hz
F1, F2 = T5.0 A
230V/50Hz
acc. to IEC 127-2
Power consumption:
800 VA
Operating
temperature range:
18°C - 35°C
Cooling system:
Refrigerant:
140 g ±5g R134A
Compressor oil:
150 ccm
EMKARATE RL15S
Dimensions:
(LxDxH)
910 x 575 x 320 mm
Weight:
50 kg
Capacities:
Paraffin reservoir:
3 l
Cassette bath:
approx.100 cass.
Heated
recessed area:
8 cassettes
Cold plate:
approx. 60 blocks
Programmable parameters:
Temperature
Paraffin reservoir
Mold warmer
Cassette bath
Work surface
Cold plate
Workdays
Daily work times (beginning, end)
Weekday
Time
Operation either manual or via the
foot pedal.
4.
Technical data
Functional unit
Temperature
Heated
Separate
Warm-up time
range
separately
temperature
prior to operation
control
Paraffin reservoir
45 - 70°C,
+
+
4 h
± 1°C increments
Mold warmer
35 - 70°C
+
+
1 h
± 1°C increments
Cassette bath
45 - 70°C
+
+
1.5 h
± 1°C increments
Work area
45 - 70°C
+
+
1 h
± 1°C increments
Cold plate
- 5°C
-
-
1 h
Forceps holder right
70°C
+
-
Forceps holder left
70°C
+
-
(not basic instrument)
Paraffin dispenser
45 - 70°C
+
-
+ pump
Temperature control
* at an ambient temperature of 22 °C and at an air humidity of 60%
All heated components include overheat protection.