11.2 Fuse replacement

Only use a fuse of the same specification! For
the required value, please refer to Chapter 4.
'Technical data' or to the fuse specification
label at the rear of the instrument.

The fuse is located at the rear of the instrument above

the left carrying handle.

Turn the fuse holder (1) a quarter turn counterclock-
wise with a screwdriver and remove.

Replace the defective fuse (2) by a new one of the
same specification.

Insert the fuse holder (1) with the fuse inserted in
the opening (3) and turn it a quarter turn clockwise
with a screw driver.

Reconnect the instrument to mains and turn it on.

11.

Maintenance

11.1 General maintenance

Only qualified and authorized service person-
nel may access the internal components of the
instrument for service and repair.

The microtome is maintenance-free in operation to a
great extent!

To ensure reliable and trouble-free operation over sev-
eral years, we recommend the following:

Have the instrument inspected by a qualified ser-
vice engineer authorized by Leica at least once a
year.

Enter into a service contract after the warranty pe-
riod; for further details, please contact your Leica
sales office.

Clean the instrument every day.

From time to time, especially after repeated drying
in a heating oven, apply a drop of cryostat oil - to
the clamping lever of the knife holder and the clamp-
ing screw for fixing the specimen discs.

Fig. 31.2

2

1

3

Clean the ventilation grids of the compressor at the
rear and side walls of the housing with a brush or
vacuum cleaner to remove dust and dirt.

Fig. 31.1