beautypg.com

7 replacing the glass anti-roll plate, Replacing the glass anti-roll plate, Daily operation – Leica Biosystems CM1100 User Manual

Page 21

background image

21

Leica CM1100 – Cryostat

I

wrong: anti-roll guide positioned too low

II

wrong: anti-roll guide positioned too high

III

right: anti-roll guide correctly positioned

In position III the anti-roll plate is correctly adjusted.

It is recommended to first preadjust it at a great sec-
tion thickness (e.g. 10 μm) and then approach the re-
quired section thickness reducing the value in 5 μm
increments and slightly readjust the anti-roll plate in
height with the knurled knob (9, Fig. 20.2).

The glass anti-roll guide is resistant to ac-
etone.
For cleaning the knife holder, domestic deter-
gents can be used.

8.

Daily operation

To raise the anti-roll plate (to move it towards the
blade), turn the knurled knob (9, Fig. 20.2) counter-
clockwise.

To lower the anti-roll plate (to retract it from the
blade), turn the knurled knob (9, Fig. 20.2) clockwise.

If the anti-roll plate is positioned incorrectly to the blade
edge, problems I and II will occur. Position III shows
the correct position of the anti-roll plate.

8.2.7 Replacing the glass anti-roll plate

Fig. 21.2

Remove the anti-roll guide from the blade by turn-
ing it to the left.

Unscrew the knurled knob (4).

Remove the white plastic washer (6) and pull the
anti-roll plate with the shaft up out of the swivel arm.

Install the new anti-roll plate:

Insert the shaft (2) of the glass anti-roll plate (1) in
the hole of the swivel arm (3) from above ensuring
the pin (6) is located in the notch (7).

Place the white plastic washer (5) on the shaft (2)
from below.

Screw the grip (4) on to the shaft (2) from below.

After installation, the anti-roll guide needs to be read-
justed as described at 8.2.6.

1

2

4

3

5

6

7

III

I

II

Fig. 21.1