Labnet Prism R Refrigerated Microcentrifuge User Manual
Page 19
16
authorized service.
Symptom
Possible Cause
Solution
Press
(START/STOP)
knob when Lid is open.
Close the lid
“
” Error
Defective
lid
lock
sensor
Call for service
Tubes are not inserted
symmetrically in rotor
holes
Load
tubes
symmetrically
Sample liquid in tubes
not properly balanced
Make sure that the
same
volume
of
liquid is in each tube
Defective or improperly
adjusted
balance
sensor
Call for service
“
” Error
(rotor
imbalance)
Rotor is stuck when a
run starts
Call for service
Lid not closed correctly
Close lid correctly
Centrifuge
cannot
be
started,
although power
is on
No speed or time has
been selected
Set speed and/or
time
No power supply
Check
power
source
Check the power
cord connection
Check the power
cord quality
Centrifuge will
not start
Blown fuse
Check the fuse and
replace if necessary