Symptom

Possible Cause

Solution

Press

(START/STOP)

knob when Lid is open.

Close the lid

” Error

Defective

lid

lock

sensor

Call for service

Tubes are not inserted

symmetrically in rotor

holes

Load

tubes

symmetrically

Sample liquid in tubes

not properly balanced

Make sure that the

same

volume

of

liquid is in each tube

Defective or improperly

adjusted

balance

sensor

Call for service

” Error

(rotor

imbalance)

Rotor is stuck when a

run starts

Call for service

Lid not closed correctly

Close lid correctly

Centrifuge

cannot

be

started,

although power

is on

No speed or time has

been selected

Set speed and/or

time

No power supply

Check

power

source

Check the power

cord connection

Check the power

cord quality

Centrifuge will

not start

Blown fuse

Check the fuse and

replace if necessary