Prism R™ Refrigerated Microcentrifuge

15

To replace rotor, first make sure the motor shaft and rotor mounting

hole are clean. Place the rotor on the motor shaft. Reinstall the rotor

securing screw on the motor shaft by turning it clockwise. Hold the

rotor with one hand and tighten the rotor securing screw, using the

rotor wrench.

6.6.

Replacing fuses

Check fuse when it is recommended in the Technical Specifications

of this manual. The fuse holder is located in the power inlet on the

right-hand side of the unit. Disconnect the power cord from the power

inlet. Open the fuse holder drawer by inserting a small screwdriver

under the tab and prying it open. Remove the innermost (operative)

fuse from its retaining tabs and replace the fuse if necessary. A spare

fuse is located in the outer most chamber of the fuse drawer. Replace

only with a fuse of the exact same value as the original.

7. Troubleshooting

To ensure the product’s quality and performance, the

Prism R

refrigerated microcentrifuge is 100% inspected by the manufacturer.

Error messages will appear on the display if the

Prism R

refrigerated

microcentrifuge fails to perform the attempted action properly. In the

case of Error (Err) messages or faults, please refer to the solutions in

the following table to clear the error messages or faults.

If the following solutions are not able to resolve the problem, please

contact Labnet International Inc.’ Service Department to arrange for