beautypg.com

Labnet Prism R Refrigerated Microcentrifuge User Manual

Page 15

background image

12

temperature can also be changed during the run.

Pre-Cooling

Pre-Cooling starts a temperature-control run at 6,000RPM.

It allows the rotor to be cooled down to the new, pre-set

temperature quickly. Cooling down from room

temperature to 4 °C takes approximately 8 min. (max. 16

minutes.) To activate Pre-Cooling function, press and hold

button, then press

(START/STOP) key. Both

RPM” and “TEMP” icons will flash during Pre-Cooling. The

Pre-Cooling

can

be

stopped

by

pressing

the

(START/STOP) key. The Pre-Cooling will function only

when the set temperature is lower than the ambient

temperature.

Standby Cool

i

ng

When the centrifuge is powered on, the refrigeration

system will cool down the chamber to the pre-set

temperature before or after the run if the chamber

temperature is above the pre-set temperature.

Note:

1. To extend the life cycle of the compressor, always remember

to inactivate the Standby Cooling (by increasing the set

temperature above room temperature) or power off the

centrifuge when it is not in use.

2. When the lid is open, the Standby Cooling will stop after 3

minutes.