beautypg.com

JUMO 707030 dTRANS T03 2-Wire/3-Wire Transmitter Data Sheet User Manual

Page 2

background image

2013-07-31/00391028

Data Sheet 707030

JUMO GmbH & Co. KG • 36035 Fulda, Germany

Page 2/10

Technical data for 2-wire transmitter
(Types 707030/..., 707031/... and 707032/...)

Input for resistance thermometer

Measurement circuit monitoring to NAMUR recommendation NE43

Output

Supply voltage

dTRANS T03 J

Type 707030/...

dTRANS T03 B

Type 707031/...

dTRANS T03 T

Type 707032/...

Measurement input

Pt100 (EN 60751)

Range limits

-200 to +850°C

Connection circuit

2-wire circuit

2-wire or 3-wire circuit

2-wire or 3-wire circuit

Smallest span

25°C

Largest span

1050°C

Unit

measuring range configuration in °C or °F

Zero shift

for spans < 75°C fixed zero: -40°C, -20°C, 0°C, 20°C, 40°C

a

a

-30 °C, -10 °C, 0 °C, 10 °C, 30 °C available upon request

for span 75°C: ±50°C

for spans > 75°C: see “Range organization” on page 7

Sensor lead resistance
for 3-wire connection

 11 per conductor

Sensor lead resistance
for 2-wire connection

factory-set:0

lead resistance

settable through setup program

Sensor current

 0.5mA

Sampling rate

continuous measurement because of analog signal path

Underrange

falling to

3.6mA

Overrange

rising to

 22mA to  28mA (typically 24mA)

Probe short-circuit

 3.6mA

Probe and lead break

positive:

 22mA to  28mA (typically 24mA)

negative:

 3.6mA

Output signal

proportional DC current 4 — 20mA

Transfer characteristic

linear with temperature

Transfer accuracy

 ± 0.1%

a

a

All details refer to the range-end value 20mA

Damping of ripple
on supply voltage

> 40dB

Burden (Rb)

Rb = (Ub - 7.5V) divided by 22mA

Burden error

 ± 0.02% per 100

a

Settling time on a temperature change

 10msec

Calibration conditions

24V DC at approx. 22°C

Calibration/configuration accuracy

 ± 0.2%

a, b, c

or

 ± 0.2°C

b

b

The larger value applies

c

If the measuring range end value > 600 °C then the calibration or configuration accuracy is

 ± 0.4 %

Supply voltage (Ub)

7.5 — 30V DC

Reverse polarity protection

yes

Supply voltage error

 ± 0.01% per V deviation from 24V

a

a

All details refer to the range-end value 20mA