JUMO 707030 dTRANS T03 2-Wire/3-Wire Transmitter Data Sheet User Manual
Brief description, Overview of function
Page 1/10
JUMO GmbH & Co. KG
Delivery address:Mackenrodtstraße 14,
36039 Fulda, Germany
Postal address: 36035 Fulda, Germany
Phone:
+49 661 6003-0
Fax:
+49 661 6003-607
e-mail:
Internet:
www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM20 2DY, UK
Phone: +44 1279 635533
Fax:
+44 1279 635262
e-mail:
Internet: www.jumo.co.uk
JUMO PROCESS CONTROL INC.
885 Fox Chase, Suite 103
Coatesville PA 19320, USA
Phone:
610-380-8002
1-800-554-JUMO
Fax:
610-380-8009
e-mail:
Internet: www.JumoUSA.com
dTRANS T03 T
Type 707032/...
dTRANS T03 TU
Type 707034/...
dTRANS T03 BU
Type 707033/...
dTRANS T03 J
Type 707030/...
dTRANS T03 B
Type 707031/...
2013-07-31/00391028
JUMO dTRANS T03 J, B, T
Analog 2-wire transmitter
with digital adjustment
JUMO dTRANS T03 BU, TU
Analog 3-wire transmitter
with digital adjustment
for connection to Pt100 resistance thermometers
for installation in: - terminal head Form B to DIN 43729
- terminal head Form J
for mounting on: - rail according to EN 60715
Brief description
These transmitters are designed for industrial applications and are used to measure the
temperature through Pt100 resistance thermometers in 2-wire or 3-wire circuit
connections (Pt500 or Pt1000 linearization upon request).
The 4
—
20 mA or 0
—
10 V output signal is linear with temperature.
The continuous analog signal path enables an extremely fast reaction time of the output
to a change in temperature (continuous analog measurement instead of digital sampling
rate), resulting in a low-noise output signal that is insensitive to interference. A very high
degree of precision - even with small ranges - is ensured thanks to the range-specific gain
adjustment.
Digital communication allows the transmitter to be adapted to the measurement task
(range, probe break and fine calibration).
Two versions are available to suit specific requirements:
Instruments with basic type extension 880/990 (adjustable)
The transmitters are calibrated for a fixed range but can, at any time, be calibrated for a
different range through the setup program.
Instruments with basic type extension 881/991 (configurable)
The required range can be configured through the setup program, without sensor
simulation and measurement.
Overview of function
dTRANS T03 J
Type 707030/...
dTRANS T03 B
Type 707031/...
dTRANS T03 T
Type 707032/...
dTRANS T03 BU
Type 707033/...
dTRANS T03 TU
Type 707034/...
Input
Pt100
Pt100
Pt100
Pt100
Pt100
Connection
circuit
2-wire
2-wire or 3-wire
2-wire or 3-wire
2-wire or 3-wire
2-wire or 3-wire
Mounting
terminal head
Form J
terminal head
Form B
mounting rail
terminal head
Form B
mounting rail
Output
4 — 20mA
4 — 20mA
4 — 20mA
0 — 10V
0 — 10V
Data Sheet 707030