12. Remote control
12.1. Communication with a PC or a superordinated data system
A suitable terminal program for communicating with a PC is:
Terminal.EXE (included with MS-Windows).
If the waterbath is put into remote control mode via the menu level,
the display will read "OFF" = REMOTE STOP. (see page 29).
The waterbath is now operated via the computer.
In general, the computer (master) sends commands to the
waterbath(slave). The waterbath sends data (including error
messages) only when the computer asks for it.
A transfer sequence consists of:
command
space
; Hex: 20)
parameter (the character separating decimals in a
group is the period)
end of file (; Hex: 0D)
The commands are divided into in or out commands.
in commands:
asking for parameters to be displayed
out commands: setting
parameters
The out commands are valid only in remote control mode.
Examples:
Command to set the working temperature T to 55.5 °C:
out_sp_00
55.5
Command to ask for the working temperature T:
in_sp_00
Response from the shaking water bath:
55.5