8.4. ATC - Absolute Temperature Calibration
Internal sensor (T
F
)
Measuring point(T
M
)
ATC serves to compensate a temperature difference that might
occur between circulator and a defined measuring point in the bath
tank because of physical properties.
The difference temperature is determined (
T= T
M
- T
T
)
and stored as correcting factor (example
T = -0.2 °C).
1.
Press the cursor key
and enter
at the same time.
2.
Use the cursor keys
to select the menu option
„At0“.
3.
With the edit keys
select "At1" and then press
enter
.
Measuring point
Enter the corrective value.
4.
Using the cursor keys
and the edit keys
set the correcting factor (example -0.20 °C) and then
press enter
.
5.
Press
and
at the same time.
The temperature on the measuring point rises to a temperature of
37.0 °C and is indicated on the
MULTI-DISPLAY (LED).
The ATC function stays activated until resetting to 00.0 °C.
Recommendation:
Use a calibrated temperature measuring instrument.