Analog inputs/outputs – JULABO HT30-M1 High Temperature Circulator User Manual
Page 50

Menu functions
50
8.5. Analog inputs/outputs
or
This submenu enables setting of the input and output values for the
programmer input and the temperature recorder outputs of socket
REG+E-PROG (31).
• Press enter
to select the inputs/outputs submenu.
• Use the up/down cursor keys
to select the desired
option and press enter
to open.
Chan.1
voltage output for recorder
(V)
Chan.2
voltage output for recorder
(V)
Chan.3
current output for recorder
(mA)
EPROG
external programmer input
First define the desired output value for channels 1 to 3:
Press the P-key
to select the desired output value and set with
enter
Setpoint
active setpoint temperature
(T, integr. programmer/ext. programmer)
ActInt internal actual temperature value
(bath temperature)
ActExt external actual temperature value
(external sensor)
Power
periodic or intermittent heating or cooling
• Then select the display size for channels 1 to 3:
Channel 1 and 2 voltage outputs
Assign the voltage values of 0 V to the lowest and 10 V to the
highest necessary temperature or power rating required as an
output value (°C or %).
Current output channel 3
Assign the current values 0 mA or 4 mA to the lowest and 20 mA to
the hightest temperature or power rating required as an output value
(°C or %).
The current output offers 2 ranges for selection:
0 to 20 mA and 4 to 20 mA.
Select the desired range by pressing the P-key
and set with enter
.
The LCD display changes automatically.