Internal / external control – JULABO HT30-M1 High Temperature Circulator User Manual
Page 36

Manual operation
36
7.5. Internal / external control
EXT
The High Temperature Circulator offers the possibility of internal
temperature control in a primary bath vessel or external control directly in
an external system.
Setup for external control:
Connect a Pt100 sensor to the socket „EXT“ of the programmable
controller, if necessary perform a calibration using the „ATC Ext:“ function
(see 8.8.
Sensors) and then securely fix the sensor in the external
system.
Setp.1: 37.00°C
ExtAct: xxxxxx°C
Power: x %
Control: extern
Setp.: 230.00°C
ExtAct: xxxxxx°C
Power: x %
Control: intern
Switching from internal to external control:
• Press the key
in operating state “OFF“ to select the control
type.
• The DIALOG-DISPLAY (LCD) indicates the effective control type.
Setp.: 37.00°C
IntAct: 26.05°C
Power: x %
Control: xxxxxx
Setp.: 37.00°C
IntAct: 25.22°C
Power: x %
Control: xxxxxx
• Press the start/stop key
.
Temperature indication:
• Both actual temperatures are indicated at the same time:
1) on the MULTI-DISPLAY (LED)
2) on the DIALOG-DISPLAY (LCD).
• Press the key
to swap the values on the displays.
The indicator light „Ext“ refers to the indication on the MULTI-
DISPLAY (LED).
Caution:
Place the external sensor into the bath medium and securely fix the sensor.