JULABO FW95-SL Ultra-Low Refrigerated Circulators User Manual
Page 88
Remote control
88
Command
Parameter
Response of circulator
in_pv_00
none
Current bath temperature.
in_pv_01
none
Heating power being used (%).
in_pv_02
none
Temperature value registered by the external Pt100 sensor.
in_pv_03
none
Temperature value registered by the safety sensor.
in_pv_04
none
Setpoint temperature (“SafeTemp”) of the excess
temperature protection
in_sp_00
none
Working temperature “Setpoint 1”
in_sp_01
none
Working temperature “Setpoint 2”
in_sp_02
none
Working temperature “Setpoint 3”
in_sp_03
none
High temperature warning limit
“OverTemp”
in_sp_04
none
Low temperature warning limit
“SubTemp”
in_sp_05
none
Setpoint temperature of the external programmer
(REG+E-PROG).
in_sp_06
none
Temperature indication in °C or °F
in_sp_07
none
1. Adjusted pump stage in the -OFF- condition.
2. Pump stage corresponding to the effective rotational
speed after start.
in_sp_08
none
Value of a flow rate sensor connected to the E-Prog input
in_sp_09
none
Value of a pressure sensor connected to the E-Prog input
in_par_00
none
Temperature difference between working sensor and safety
sensor
in_par_01
none
Te - Time constant of the external bath.
in_par_02
none
Si - Internal slope
in_par_03
none
Ti - Time constant of the internal bath.
in_par_04
none
CoSpeed - Band limit (max. difference between the
temperatures in the internal bath and external system).
in_par_05
none
Factor pk/ph0:
Ratio of max. cooling capacity to max. heating capacity
in_par_06
none
Xp control parameter of the internal controller.
in_par_07
none
Tn control parameter of the internal controller.
in_par_08
none
Tv control parameter of the internal controller.
in_par_09
none
Xp control parameter of the cascade controller.
in_par_10
none
Proportional share of the cascade controller.
in_par_11
none
Tn control parameter of the cascade controller.
in_par_12
none
Tv control parameter of the cascade controller.
in_par_13
none
Adjusted maximum internal temperature of the cascade
controller.
in_par_14
none
Adjusted minimum internal temperature of the cascade
controller.
in_par_15
none
Band limit (upper)
in_par_16
none
Band limit (lower)