List of commands – JULABO FW95-SL Ultra-Low Refrigerated Circulators User Manual
Page 86
Remote control
86
The out commands are valid only in remote control mode.
Examples:
When the RS485 interface is used, the three-digit instrument address
precedes each command.
(example: address Ad32 = A0 32 )
Command to set the working temperature >Setpoint1< to 55.5 °C
out_sp_00
⇔ 55.5↵
A032_out_sp_00
⇔ 55.5↵
Command to retrieve the working temperature >Setpoint1<
in_sp_00
↵
A032_in_sp_00
↵
Response from the circulator:
55.5
↵ LF
A032_55.5
↵ LF
Notice:
out commands
Temperature values can be displayed on the circulator in °C or °F.
The settings sent via the interface must also be made in the selected unit according
to this setting.
12.3. List of commands
out commands: Setting parameters or temperature values.
Command
Parameter
Response of circulator
out_mode_01
0
Use working temperature >Setpoint1<
out_mode_01
1
Use working temperature >Setpoint2<
out_mode_01
2
Use working temperature >Setpoint3<
out_mode_02
0
Selftuning “off”.
Temperature control using the stored parameters.
out_mode_02
1
Selftuning “once”
Single selftuning of the controlled system after the next start.
out_mode_02
2
Selftuning “always”
Continual selftuning of the controlled system whenever a new
setpoint is to be reached.
out_mode_03
0
Set external programmer input to voltage.
Voltage
0 V ... 10 V
out_mode_03
1
Set external programmer input to current.
Current
0 mA ... 20 mA
out_mode_04
0
Temperature control of internal bath.
out_mode_04
1
External control with Pt100 sensor.
out_mode_05
0
Stop the unit = R –OFF-.