Limits for internal control and external control, E 67 – JULABO FW95-SL Ultra-Low Refrigerated Circulators User Manual

Page 67

background image

HL, SL

67

The parameter flashes. Set by using the
numeric keypad and confirm by pressing

.

The parameter flashes. Set by using the
numeric keypad and confirm by pressing

.

The parameter flashes. Set by using the
numeric keypad and confirm by pressing

.

The parameter flashes. Set by using the
numeric keypad and confirm by pressing

.

9.8.1.

Limits for internal control and external control

Set Max / Min– Maximum and minimum setpoint
Restriction of the adjustable temperature range.

The limitation of the operating temperature range effects the temperature

setting in the menu with the key

.

Only setting of working temperatures which lie within the determined limits
is possible
Existing settings for Setpoint 1, -2, -3, as well as those for >OverTemp<
and > SubTemp <, are automatically deferred into the limit range.

Setting range: -94,90 °C ... +300,0 °C

Set maximum heating / cooling

The heating and cooling capacity of the unit are adjustable. 100 %
corresponds to the technical specification of the equipment.

Setting range:

Heating Max –

0 to 100 % in 1 % steps

Cooling Max –

0 to 100 % in 1 % steps