Setting of temperatures, Safety installations, warning functions – JULABO FW95-SL Ultra-Low Refrigerated Circulators User Manual

Page 29

background image

HL, SL

29

7.

Setting of temperatures

Press the

key to call up the menu for temperature selection.

3 different working temperatures can be adjusted. Their values are freely
selectable within the operating temperature range.

The temperatures can be set in start or stop mode.

Setting of working temperature in the

menu

Factory settings:

1. Press the key

. The value flashes

2. Select SETPOINT 1 or 2 or 3 using the key

or

.

3. Confirm by pressing the

key.

The circulator uses the new working temperature value for
temperature control.

Example:
Adjustment/modification of the pre-setting of "SETPOINT 3"

If the active Setpoint is changed, the new value is immediately used

for the control of the working temperature.

The heater control indicator flashes.

Setpoint3:

Change the value from 70.00°C to 80.00°C.

See standard display line 1

• Use the numeric keypad to enter

and

.

Then press enter

to store the value.

Notice: Refer to SETPOINT MAX / MIN in chapter „LIMITS“






8.

Safety installations, warning functions

Check the safety installations at least twice a year! Refer to page 17.

Settings for the excess temperature protection > SafeTemp< and for the

warning functions > OverTemp< and > SubTemp< are made in a menu

which is called up by pressing the key

.

Menu item > Type< allows choosing between a warning and an
alarm cut-off for the menu items > OverTemp< and > SubTemp<.