beautypg.com

Setting the pump pressure, E 45 – JULABO FW95-SL Ultra-Low Refrigerated Circulators User Manual

Page 45

background image

HL, SL

45

9.3. Setting the pump pressure


The capacity of the circulating pump is set by adjusting the motor speed

Press the

key if a parameter is to be retained. (prior to OK)

Level 2

Submenu >Pump<

The set pump stage is displayed. Example: 1

• Confirm by pressing

.

• Set required pump stage by pressing

and

.

Example: 3


Example:

Adjusted

Effective

Adjustable pump capacity

stage 1 ... 4

Illuminated display:

for pump pressure

Indication of the adjusted pump pressure stage in the –OFF- mode.

Indication of the effective pump pressure stage (rotational speed)
after start.

To protect the pump motor, the rotational speed changes with the load
applied, for example, depending on the viscosity of the bath fluid at
different working temperatures.

Flow rate:

22 ... 26 l/min

Pump capacity stage

1

2

3

4

Pump pressure

[bar]

0.4

0.5

0.6

0.7

Suction pump

[bar]

0.2

0.26

0.33

0.4

Total capacity

[bar]

0.6

0.76

0.93

1.1

in a loop circuit