JULABO FW95-SL Ultra-Low Refrigerated Circulators User Manual
Page 87
HL, SL
87
Command
Parameter
Response of circulator
out_mode_05
1
Start the unit.
out_mode_08
0
Set the control dynamics - aperiodic
out_mode_08
1
Set the control dynamics - standard
out_sp_00
xxx.xx
Set working temperature. “Setpoint 1”
out_sp_01
xxx.xx
Set working temperature. “Setpoint 2”
out_sp_02
xxx.xx
Set working temperature. “Setpoint 3”
out_sp_03
xxx.xx
Set high temperature warning limit “OverTemp”
out_sp_04
xxx.xx
Set low temperature warning limit “SubTemp”
out_sp_06
xxx.xx
Set manipulated variable for heater via serial interface
-100 ... +100 [%]
out_sp_07
x
Set the pump pressure stage. (1 ... 4)
out_par_04
x.x
CoSpeed for external control 0 ... 5.0.
out_par_06
xxx
Xp control parameter of the internal controller.
0.1 ... 99.9
out_par_07
xxx
Tn control parameter of the internal controller.
3 ... 9999
out_par_08
xxx
Tv control parameter of the internal controller.
0 ... 999
out_par_09
xxx
Xp control parameter of the cascade controller.
0.1 ... 99.9
out_par_10
xxx
Proportional share of the cascade controller.
1 ... 99.9
out_par_11
xxx
Tn control parameter of the cascade controller.
3 ... 9999
out_par_12
xxx
Tv control parameter of the cascade controller.
0 ... 999
out_par_13
xxx
Maximum internal temperature of the cascade controller.
out_par_14
xxx
Minimum internal temperature of the cascade controller.
out_par_15
xxx
Band limit (upper)
0 ... 200 °C
out_par_16
xxx
Band limit (lower)
0 ... 200 °C
out_hil_00
-xxx
Set the desired maximum cooling power (0% to 100%).
Note: Enter the value with a preceding negative sign!
This setting is meaningful only for FP cooling machines.
out_hil_01
xxx
Set the desired maximum heating power (10% to 100%).
in commands: Asking for parameters or temperature values to be displayed.
Command
Parameter
Response of circulator
version
None
Version number of the software (V X.xx)
status
none
Status message, error message (see page 89)