JULABO FW95-SL Ultra-Low Refrigerated Circulators User Manual
Page 12

Operator responsibility – Safety recommendations
12
HL
SL
Temperature selection
digital
via keypad
indication on LCD DIALOG-DISPLAY (°C/°F)
remote control via PC
indication on monitor
Temperature indication
VFD COMFORT-DISPLAY (°C/°F)
Resolution
°C
0.01
ATC3
INT / EXT
°C
±3 / ±9
Temperature control
ICC - Intelligent Cascade Control
Heater wattage (at 230 V)
Heater wattage (at 115 V)
kW
2.0
3.0
Temperature selection
kW
1.0
-----
Electronically adj. pump capacity
stages
1 ... 4
1 ... 4
Flow rate
at 0 bar
l/min
22 ... 26
22 ... 26
Max. pressure
at 0 liters
bar
0.7
0.7
Max. suction
at 0 liters
bar
0.4
0.4
Electrical connections:
See page 13
Ambient temperature
°C
5 ... 40
5 ... 40
Mains power connection 230 V/50-60 Hz
V/ Hz
190 ... 253 / 50-60
190 ... 253 / 50-60
Current draw
(at 208 V / 230 V)
A
8 / 9
12 / 13
Mains power connection 115 V/50-60 Hz
V/ Hz
90 ... 127 / 50-60
-----
Current draw
(at 100 V / 115 V)
A
9/10
-----
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C
Technical changes without prior notification reserved.