JULABO FW95-SL Ultra-Low Refrigerated Circulators User Manual
Page 61

HL, SL
61
Press the
key if a parameter is to be retained. (prior to OK)
Level 2
Parameter-Level
or
•
The parameter flashes, switch by pressing
and
or
The parameter flashes, switch by
pressing
and
>no< Carry out an ATC calibration
>yes< return to standard operation after
calibration.
•
The parameter flashes, switch by pressing
and
A >1-point<, >2-point< or >3-point<
calibration can be carried out.
The value >TmpValue< is only indicated
In addition the measured temperature
value >CalValue X< is saved during the
next step.
•
The value flashes. Set by using the
numeric keypad and confirm by pressing
.
If only a 1-point calibration is carried out, the following menu items are
not indicated anymore