9.1.5.

Interruption after a power failure









It is possible to operate the programmer safely in case of a power
failure.
Important:

For this, >Autostart< needs to be set in the >Configuration< menu.

L

If the AUTOSTART function is activated, the programmer starts
again at a point approx. 20 seconds before the interruption took
place. However, an uncontrolled change of the bath temperature has
occurred.

L

Power failure when the instrument is on hold if start is effected via
the built-in timer:
If the starting time is not yet reached, the programmer is put on hold
again.
If the starting time is exceeded, the programmer starts immediately.


9.1.6.

Termination of a program

L

A program can be terminated by pressing

. The programmer

switches back to the Start menu.

Further procedures are defined in the MENU >START PROGRAM<,
menu item >Program End <.

Example: >Program End - StdBy<

The VFD COMFORT-DISPLAY displays “OFF”.



Interruption after a power failure

In case of a power failure, the reaction of the circulator is the same as
when switched off and on again with the mains switch.

The VFD COMFORT-DISPLAY displays “OFF”.