beautypg.com

Start - stop – JULABO FPW91-SL Ultra-Low Refrigerated Circulators User Manual

Page 33

background image

SL

33

Selecting the language:
There are two options for the language of the LCD DIALOG-DISPLAY:
German or English. Select the desired language in the menu
>Configuration< under the submenu >Language/Sprache<.
Press the respective keys in the following order:

3x up to menu Configuration

6x up to submenu >Language/Sprache<

deutsch / english


6.2.2.

Switching on the cooling machine


Switching on:

• Switch on the cooling machine using the switch (1b) .

L

Control of the cooling machine

With the mains switch (1b) turned on, the circulator automatically
switches the cooling machine off and on.
• It is switched off if:

- the current working temperature is increased by > 30 °C (cooling is
not required).
- the heater operates at full power (> 800 W) for longer than 5
minutes.

• It is switched on if:

- cooling is necessary to maintain the bath temperature.
After switch-off, the cooling machine automatically switches on only
after a delay of 5 minutes to protect the cooling compressor.

L

To save energy, turn off the cooling machine with the mains switch
(1b) whenever cooling is not required.


6.3. Start - Stop

Start:

• Press

key.

The actual bath temperature is displayed on the VFD COMFORT-
DISPLAY. The circulating pump starts with a slight delay.

Stop:

• Press

key.

or
Keep

key pressed.

The VFD COMFORT-DISPLAY indicates the message "OFF".