beautypg.com

JULABO FL601 Recirculating Coolers User Manual

Page 41

background image

41

11.4. Status

messages

Status messages

Description

00 MANUAL STOP

Recirculating cooler in „OFF“ state.

01 MANUAL START

Recirculating cooler in keypad control mode.

02 REMOTE STOP

Recirculating cooler in „r OFF“ state.

03 REMOTE START

Recirculating cooler in remote control mode.

11.5. Error

messages

Error messages

Description

-01 LOW LEVEL ALARM

Low liquid level alarm.

-05 WORKING SENSOR ALARM

Working temperature sensor short-circuited or
interrupted.

-03 EXCESS TEMPERATURE

WARNING

High temperature warning. Starting at 75 °C
(no deactivation)
The return temperature soon reaches the switch-off
value of the high temperature warning function (85 °C)

-07 I

2

C-BUS ERROR

Internal error when reading or writing the I

2

C bus.

-08 INVALID COMMAND

Invalid command.

-09 COMMAND NOT ALLOWED IN

CURRENT OPERATING MODE

Invalid command in current operating mode.

-10 VALUE TOO SMALL

Entered value too small.

-11 VALUE TOO LARGE

Entered value too large.

This manual is related to the following products: