Important times for a command transmission:

To ensure a safe data transfer, the time gap between two commands should
be at least 250 ms.
The recirculating cooler automatically responds to an in command with a data
string followed by a LF (Line Feed). The next command should only be sent
after 10 ms.

The commands are divided into in or out commands.

in commands:

asking for parameters to be displayed

out commands: setting

parameters

The out commands are valid only in remote control mode.


Examples:
Command to set the working temperature to15,5 °C:
out_sp_00

15.5

Command to ask for the working temperature
in_sp_00

Response from the recirculating cooler:
15.5

LF