11.3. List of commands

out commands: Setting parameters or temperature values.

Command

Parameter

Response of recirculating cooler

out_mode_05

0

Stop the unit = R –OFF-.

out_mode_05

1

Start the unit.

out_sp_00 xxx.xx Set working temperature

in commands: Asking for parameters or temperature values to be displayed.

Command

Parameter

Response of recirculating cooler

version none

Number

of

software version (V X.xx)

status none

Status message, error message (see page 41 )

in_pv_00

none

Actual bath temperature.

in_sp_00 none Working temperature

in_mode_05 none

Recirculating cooler in Stop/Start condition:
0 = Stop
1 = Start