Alarm output

Potential-free change-over contact for external alarm signal.

Pin 2 and 3 are connected in case of an alarm.
Pin 2 and 1 are connected in normal condition or mains switch "Off".

or


Switching capacity max. 30 W / 30 VA
Switching voltage max. 30 V

/

Switching current max. 1 A

11. Remote

control

11.1. Setup for remote control

9

6

5

1

RS232

Check the interface parameters for both interfaces (on recirculating cooler and
PC) and make sure they match.

Interface parameters are pre-determined.

Type RS232

Baudrate 4800

bauds

Parity even
Handshake hardware

handshake