Page 24

background image

24

tina49e1-c (2009-06)

Push the button again:

Fine adjustment
within 0...3 s:

the zero adjustment value changes
by one unit

Change of direction
within 3...5 s:

the zero adjustment changes its
direction (the flashing frequency of
the LED changes briefly)

If the button is released for more than 5 s, the

gauge returns to the measurement mode.

The zero with Base-Pressure-Offset can also be ad-

justed via the RS232C interface (

→ [4]).

The LED flashes if the signal output is negative.

4.3 Switching

Functions

(CDG025D-S only)

The two switching functions can be adjusted to any pressure
within the whole measurement range (

→ 13).

The current setpoint setting
• is output at the D-Sub connector instead of the measurement

signal (

→ 18) and can be measured with a voltmeter after

the button is pressed, or

• can be read/written via the RS232C interface.
If the pressure is lower than the setpoint, the corresponding LED
is lit (<1> or <2>) and the corresponding relay (

→ 18) is

energized.