INFICON Cube CDGsci (RS232C) User Manual
Page 9

tira90e1 (2014-01) RS232C.cp
9
Parameter name
Data
type
Access
Right
Byte
Type
Byte
No. 1
Byte
No. 2
Byte
No. 3
Comment
SWDateMonthDay uint16 R H-Byte
0x00
0xD6
Software version date Month in Hex
e.g. 0x03 = March
L-Byte
0xD7
Software version date Day in Hex
e.g. 0x19 = 19
Part No.
uint8
R
Byte 0
0x00
0xD8
Part number as ASCII string
uint8
R
Byte 1
0x00
0xD9
(Max 20 byte)
uint8
R
…
0x00
…
e.g.
378-000
uint8
R
Byte 19
0x00
0xE8
(Last digit: null terminator)
RemainingZero
sint16
R
H-Byte
0x00
0x48
Max. remaining offset value
L-Byte
0x49
ExtendedError
uint8
R
H-Byte
0x00
0x36
Bit 0
PT1000 fault
H-Byte
Bit 1
Heater block overtemp.
Bit 2
Electronic overtemp.
Bit 3
Zero adjust error
Bit
4
Reserve
Bit
5
Reserve
Bit
6
Reserve
Bit
7
Reserve
ExtendedError
uint8
R
L-Byte
0x00
0x37
Bit 0
Atm. pressure out of range
L-Byte
Bit 1
Temperature out of range
Bit
2
Reserve
Bit
3
Reserve
Bit 4
Cal. mode wrong
Bit 5
Pressure underflow
Bit 6
Pressure overflow
Bit 7
Zero adjust warning
SensPressRange uint8 R
0x00 0x38
0 F.S.R.
=
E-3
(Exponent)
1
F.S.R.
=
E-2
2
F.S.R.
=
E-1
3
F.S.R. = E 0
4
F.S.R.
=
E+1
5
F.S.R.
=
E+2
6
F.S.R.
=
E+3
7
F.S.R.
=
E+4
SensFSR
uint8
R
0x00
0x39
0
Mantissa = 1.0
(Mantissa)
1
Mantissa = 1.1
2
Mantissa = 2.0
3
Mantissa = 2.5
4
Mantissa = 5.0
FirmwareRevisionCPU2
uint8
R
Byte 0
0x00
0x10
Software version Cube application
program
uint8
R
Byte
1
0x00
uint8
R
…
x
e.g. 1.6.0.14
uint8
R
Byte 19
x
(Last digit: null terminator)
ImageRevisionCPU2 uint8
R
Byte
0 0xE0 0x24
Image version of WinCE operating
system
uint8
R
Byte
1
0xE0
uint8
R
…
0xE0
e.g.
WinCE60
uint8
R
Byte 19
0xE0
(Last digit: null terminator)
(continued)