2 Parameter
Table
All values in this table are quoted in HEX, thus the prefix '0x'.
Parameter name
Data
type
Access
Right
Byte
Type
Byte
No. 1
Byte
No. 2
Byte
No. 3
Comment
DataTxMode uint8
R
0x00
0x00
0x00
Continued output of measured value
W
0x10
0x01
Individual output of measured value
(polling)
CPU2Unit
uint8
R
0x00
0x01
0x00
Pressure unit "mbar"
W
0x10
0x01
Pressure unit "Torr"
0x02
Pressure unit "Pa"
Filter uint8
R
0x00
0x02 0x00
Filter
dynamic
W
0x10
0x01
Filter time fast
0x02
Filter time slow
SP1 Level Low
sint16
R
H-Byte
0x00
0x04
Lower setpoint threshold SP1
1)
W
H-Byte
0x10
L-Byte
0x05
SP2 Level Low
sint16
R
H-Byte
0x00
0x06
Lower setpoint threshold SP2
1)
W
0x10
L-Byte
0x07
SP1 Level High
sint16
R
H-Byte
0x00
0x08
W
0x10
L-Byte
0x09
Upper setpoint threshold SP1
(hysteresis)
1)
SP2 Level High
sint16
R
H-Byte
0x00
0x0A
W
0x10
Upper setpoint threshold SP2
(hysteresis)
1)
L-Byte
0x0B
SP1, percentage of ATM (only for
1000Torr)
PerOfAtmSP1 uint8
R
0x00
0xFC
W
0x10
SP2, percentage of ATM (only for
1000Torr)
PerOfAtmSp2 uint8
R
0x00
0xFD
W
0x10
FirmwareRevisionCPU1
uint8
R
0x00
0x10
(
CPU1
e.g. 20 = V1.0
Calib date
uint32
R
MSB
0x00
0x11
Date:
YY,MM,DD,HH,MM
R
Byte 2
0x00
0x12
e.g. 0410291109 =
R
Byte 1
0x00
0x13
2004-10-29 at 11:09
R
LSB
0x00
0x14
ZeroAdjustValue
sint16
R
H-Byte
0x00
0x15
Zero Offset Adjust Level
2)
W
0x10
L-Byte
0x16
DCOutputOffset sint16
R
H-Byte
0x00
0x17
Customer DC Output Offset
2)
W
0x10
L-Byte
0x18
(Base pressure offset)
SerialNumber
uint8
R
Byte 0
0x00
0x19
Production number as ASCII string
(barcode)
uint8
R
Byte 1
0x00
0x1A
(Max. 16 byte)
uint8
R
…
0x00
…
...
uint8
R
Byte 15
0x00
0x28
(Last digit: null terminator)
SWDateYear
uint16
R
H-Byte
0x00
0xD4
Software version date Year in Hex
e.g. 0x2007 = 2007
L-Byte
0xD5
(continued)