INFICON Cube CDGsci (RS232C) User Manual
Page 13

tira90e1 (2014-01) RS232C.cp
13
The "DC Output Level" variable is used for assigning a defined offset level to the
analog output signal, e.g. for setting a certain zero offset signal level.
A "DC Output Level" >0 reduces the output range of the measurement
range 0 … 10 V by the selected offset value (F.S.R. - DC output level).
The "DC Output Level" parameter (16-Bit) consists of the high and low byte.
DC Output Level =
b
×F.S.R._Mantisse×10
(F.S.R._Exp)
Parameter
1)
Description
DC Output Level DC-Output-Signal in the selected pressure unit
(
→ 13, table "Parameter").
1)
Further
parameter
→ 13, table "Parameter"
Software version =
20
e.g.
≙ V1.0
Parameter Description
Parameter Description
byte" (16 bit value), data value in decimal format.
a Conversion factor for pressure units other than "Torr"
Torr:
a = 1.00
mbar:
a = 1.3332
Pa:
a = 133.32
b Factor for resolution
Page
No.
1)
b Output
signal
2
8’192’000
10.20
V
3
8’192’000
10.20
V
4
8’388’352
10.00
V
F.S.R._Mantisse F.S.R.
factor
according to the "Sensor type" variable, which
has to be read separately (
→ "Read command").
F.S.R._Exp F.S.R. exponent according to "Sensor type" variable, which
has to be read separately (
→ "Read command").
1)
→ 4, tabel "Structure of send string", byte no. 1
DC Output Level
Software version