INFICON Edge CDG045D2 1 … 1000Torr / mbar User Manual
Page 17
tina86e1 (2013-11)
17
DANGER
DANGER: overpressure in the vacuum system
>2.5 bar
KF flange connections with elastomer seals (e.g.
O-rings) cannot withstand such pressures. Process
media can thus leak and possibly damage your
health.
Use O-rings provided with an outer centering ring.
DANGER
DANGER: protective ground
Products that are not correctly connected to ground
can be extremely hazardous in the event of a fault.
Electrically connect the gauge to the grounded
vacuum chamber. This connection must conform to
the requirements of a protective connection ac-
cording to EN 61010:
• CF and VCR flanges fulfill this requirement.
• For gauges with a KF flange, use a conductive
metallic clamping ring.
• For gauges with a ½" tube, take appropriate
measures to fulfill this requirement.
Caution
Caution: vacuum component
Dirt and damages impair the function of the vac-
uum component.
When handling vacuum components, take appro-
priate measures to ensure cleanliness and prevent
damages.