INFICON SKY CDGxxxD (RS232C) User Manual
Page 9

tira49e1-a (2010-02) RS232C.cp
9
(completed)
Parameter name
Data type
Description
Byte No. 2 Byte No. 3 Comment
Extended error
uint8_T / Ro
H-Byte
54
Bit 0
PT1000 fault
4)
H-Byte
Bit 1
Heater block overtemp.
4)
Bit 2
Electronic overtemp.
4)
Bit 3
Zero adjust error
Bit
4
Reserve
Bit
5
Reserve
Bit
6
Reserve
Bit
7
Reserve
Extended error
uint8_T / Ro
L-Byte
55
Bit 0
Atm. pressure out of range
L-Byte
Bit 1
Temperature out of range
Bit
2
Reserve
Bit
3
Reserve
Bit 4
Cal. mode wrong
Bit 5
Pressure underflow
Bit 6
Pressure overflow
Bit 7
Zero adjust warning
Pressure range
uint8_T / Ro
56
0
F.S.R. = E-3
(Exponent)
1
F.S.R.
=
E-2
2
F.S.R.
=
E-1
3
F.S.R. = E 0
4
F.S.R.
=
E+1
5
F.S.R.
=
E+2
6
F.S.R.
=
E+3
7
F.S.R.
=
E+4
Pressure range
uint8_T / Ro
57
0
Mantissa = 1.0
(Mantissa)
1
Mantissa = 1.1
2
Mantissa = 2.0
3
Mantissa = 2.5
4
Mantissa = 5.0
Gauge config
uint8_T / Ro
58
0
= Analog out 0 … 10.24 V
1
= Analog out 1 … 9 V
CDG type
uint8_T / Ro
59
0
= CDG025D
1
=
CDG045D
2
=
CDG100D
3
=
CDG160D
4
=
CDG200D
RW = Read / Write
Ro = Read only
1)
Factory
setting
2)
Conversion
→ Section 1.1, byte 6 "Read command"
3)
Conversion
→ Section 1.1, bytes 4 and 5 "Pressure unit"
4)
CDG045D, CDG100D, CDG160D, CDG200D only