INFICON SKY CDGxxxD (RS232C) User Manual

Page 4

background image

4

tira49e1-a (2010-02) RS232C.cp

Bit No. Definition

0 RS232

synchronization

error

1

Incorrect command, e.g. inadmissible address (syntax error)

2

Inadmissible read command

3 SP1

status

4 SP2

status

5 Not

used

6 Not

used

7

Extended error set (

→ Read command "Extended Error

L-Byte and H-Byte")

No bit set

→ value = 0x00 = no error set

Errors are only recorded in the error bytes as long as they persist, except for
RS232 interface errors. Errors are not acknowledged.
RS232 errors are signaled by the "toggle bit", i.e. when an RS232 error occurs, the
"toggle bit" is not inverted. For checking the status of the "toggle bit", a read
operation is required, which also allows to read the error byte for detailed error
analysis.
If an "extended error" is set, it has to be read as variable by means of the "read
command" (

→ table "Variables for bytes No. 2 and 3"). After the read operation, the

variable is automatically erased.

The pressure is calculated from bytes 4 and 5 of the send string (decimal
presentation).

×a

p =

b

×F.S.R._Mantissa×10

(F.S.R._Exp)


Parameter Description

p Pressure value in selected pressure unit (→ Parameter a)

Pressure measurement data, composed of "low and high

byte" (16 bit value) and converted into decimal format

F.S.R._Mantissa F.S.R.

factor

according to "Sensor type" variable, which

has to be read separately (

→ "Read command")

F.S.R._Exp F.S.R. exponent according to "Sensor type" variable, which

has to be read separately (

→ "Read command")

a Conversion factor for pressure units other than "Torr".

Torr:

a = 1.00

mbar:

a = 1.3332

Pa:

a = 133.32

b Factor for resolution

Page

No.

1)

b Output

signal

2

32000

10.24

V

3

32000

10.24

V

4

32767

10.00

V

1)

→ 3, tabel "Structure of send string", byte no. 1


Error byte
(byte No. 3)

Error handling

Calculation of pressure value

Conversion formula
(byte No. 4 and 5)