tina60e1-a (2011-04)

45

4.5 DeviceNet

Operation

Caution

Caution: data transmission errors
The attempt to operate the DeviceNet gauge with
the RS232C interface causes data transmission
errors.
This DeviceNet gauge must not be operated with
the RS232C interface.

Before the gauge is put into operation, it has to be configured for
the DeviceNet. A configuration tool and the device specific EDS
(Electronic Data Sheet) file are required for this purpose. This
software can be downloaded from our website.

Node Address Setting

Set the node address (0 … 63

dec

) via the

, , and switches
(default 63

dec

). The node address is polled by

the firmware when the gauge is switched on.
If the setting deviates from the stored value,
the new value is taken over into the NVRAM.
If a setting higher than 63 is made, the pre-
vious node address setting remains valid.
If the switch is in the

position,
the node address is programmable via the
DeviceNet (

→ [4]).