21

Valid in the range 0.375 … 2.842 V

p = a + bU + cU

2

+ dU

3

+ eU

4

+ fU

5

a –0.02585 c 0.04563 e –0.04158
b 0.03767

d 0.1151

f 0.008737

where p pressure in Torr

a, b, c, d, e, f constant

U output signal


Valid in the range 2.842 … 4.945 V

a + cU + eU

2

p =

1 + bU + dU

2

+ fU

3

a 0.1031 c –0.02322 e 0.07229
b –0.3986 d 0.07438 f –0.006866

where p pressure in Torr

a, b, c, d, e, f constant

U output signal


Valid in the range 4.945… 5.529 V

a + cU

p =

1 + bU + dU

2

a 100.624 c –20.5623
b –0.37679 d 0.0348656

where p pressure in Torr

a, b, c, d constant