Data format, Examples, Fixs32enxx – INFICON PSG55x ATM to Medium Vacuum Gauge (RS232C / RS485C) User Manual
Page 9
tira59e1 (2010-07) RS232C.cp
9
(concluded)
PID
Name
Description
Factory
setting
Min.
Max. Type
457
High Trip Point 1 Hysteresis
Hysteresis High Trip Point 1 [mbar]
1.00E+01
5.00E-05
1500 Fixs32en20
458
Low Trip Point 1 Hysteresis
Hysteresis Low Trip Point 1 [mbar]
5.00E-05
5.00E-05
1500 Fixs32en20
459
High Trip Point 2 Hysteresis
Hysteresis High Trip Point 2 [mbar]
1.00E+01
5.00E-05
1500 Fixs32en20
460
Low Trip Point 2 Hysteresis
Hysteresis Low Trip Point 2 [mbar]
5.00E-05
5.00E-05
1500 Fixs32en20
461
Setpoint 1 Extended Status
Extended status of Setpoint 1
0 Not
active
1
Low Trip Point active
2
High Trip Point active
3
High and Low Trip Point active
0
Uint8
462
Setpoint 2 Extended Status
Extended status of Setpoint 2
0 Not
active
1
Low Trip Point active
2
High Trip Point active
3
High and Low Trip Point active
0
Uint8
To calculate the result from a Fixx32enXX value, multiply or divide it by a factor of
2
XX
.
Example Fixs32en20
The hysteresis of the High Trip Point Setpoint 1 is to be set to 10 mbar. For this
purpose 10 is multiplied by 2
20
(Fixs32en20). The value 10485760 is therefore
transmitted.
The following examples show how read and write requests are made.
The required parameter has PID 221.
From Master to PCG5xx:
0 1 2 3 4 5 7 9 9+n
MSB
LSB
MSB
LSB
Address
Device ID
Ack
MsgLength
Cmd
PID
Reserved
Data
Crc
0x00
0x00
0x00
0x05
0x01
0x00DD
0x0000
-
0xAB21
From PCG5xx to Master:
0 1 2 3 4 5 7 9 9+n
MSB
LSB
MSB
LSB
Address
Device ID
Ack
MsgLength
Cmd
PID
Reserved
Data
Crc
0x00
0x02
0x01
0x09
0x02
0x00DD
0x0000
0x375A05BF
0xD9BB
The required parameter has PID 224. To set the unit to Torr, 1 must be written.
From Master to PCG5xx:
0 1 2 3 4 5 7 9 9+n
MSB
LSB
MSB
LSB
Address
Device ID
Ack
MsgLength
Cmd
PID
Reserved
Data
Crc
0x00
0x00
0x00
0x06
0x03
0x00E0
0x0000
0x01
0x346D
From PCG5xx to Master:
0 1 2 3 4 5 7 9 9+n
MSB
LSB
MSB
LSB
Address
Device ID
Ack
MsgLength
Cmd
PID
Reserved
Data
Crc
0x00
0x02
0x01
0x05
0x04
0x00E0
0x0000
-
0x94EA
Data Format
Fixs32enXX
Examples
Read request (reading a
pressure)
Write request (setting a unit)