Data format, Examples, Fixs32enxx – INFICON PSG55x ATM to Medium Vacuum Gauge (RS232C / RS485C) User Manual

Page 9

background image

tira59e1 (2010-07) RS232C.cp

9

(concluded)

PID

Name

Description

Factory

setting

Min.

Max. Type

457

High Trip Point 1 Hysteresis

Hysteresis High Trip Point 1 [mbar]

1.00E+01

5.00E-05

1500 Fixs32en20

458

Low Trip Point 1 Hysteresis

Hysteresis Low Trip Point 1 [mbar]

5.00E-05

5.00E-05

1500 Fixs32en20

459

High Trip Point 2 Hysteresis

Hysteresis High Trip Point 2 [mbar]

1.00E+01

5.00E-05

1500 Fixs32en20

460

Low Trip Point 2 Hysteresis

Hysteresis Low Trip Point 2 [mbar]

5.00E-05

5.00E-05

1500 Fixs32en20

461

Setpoint 1 Extended Status

Extended status of Setpoint 1

0 Not

active

1

Low Trip Point active

2

High Trip Point active

3

High and Low Trip Point active

0

Uint8

462

Setpoint 2 Extended Status

Extended status of Setpoint 2

0 Not

active

1

Low Trip Point active

2

High Trip Point active

3

High and Low Trip Point active

0

Uint8

To calculate the result from a Fixx32enXX value, multiply or divide it by a factor of
2

XX

.

Example Fixs32en20

The hysteresis of the High Trip Point Setpoint 1 is to be set to 10 mbar. For this
purpose 10 is multiplied by 2

20

(Fixs32en20). The value 10485760 is therefore

transmitted.

The following examples show how read and write requests are made.

The required parameter has PID 221.

From Master to PCG5xx:

0 1 2 3 4 5 7 9 9+n

MSB

LSB

MSB

LSB

Address

Device ID

Ack

MsgLength

Cmd

PID

Reserved

Data

Crc

0x00

0x00

0x00

0x05

0x01

0x00DD

0x0000

-

0xAB21

From PCG5xx to Master:

0 1 2 3 4 5 7 9 9+n

MSB

LSB

MSB

LSB

Address

Device ID

Ack

MsgLength

Cmd

PID

Reserved

Data

Crc

0x00

0x02

0x01

0x09

0x02

0x00DD

0x0000

0x375A05BF

0xD9BB

The required parameter has PID 224. To set the unit to Torr, 1 must be written.

From Master to PCG5xx:

0 1 2 3 4 5 7 9 9+n

MSB

LSB

MSB

LSB

Address

Device ID

Ack

MsgLength

Cmd

PID

Reserved

Data

Crc

0x00

0x00

0x00

0x06

0x03

0x00E0

0x0000

0x01

0x346D

From PCG5xx to Master:

0 1 2 3 4 5 7 9 9+n

MSB

LSB

MSB

LSB

Address

Device ID

Ack

MsgLength

Cmd

PID

Reserved

Data

Crc

0x00

0x02

0x01

0x05

0x04

0x00E0

0x0000

-

0x94EA

Data Format

Fixs32enXX

Examples

Read request (reading a
pressure)

Write request (setting a unit)