Protocol description, Communication error, Output pressure error – INFICON PSG55x ATM to Medium Vacuum Gauge (RS232C / RS485C) User Manual
Page 5

tira59e1 (2010-07) RS232C.cp
5
If a communication error occurs during transmission the PID is set to 0xFFFF and
an error byte is added.
Byte 0, errors
Error Code Description
1 Access
error
2
Value higher max. or lower min.
3
Parameter not found
4 Length
error
6 Memory
access
error
7 Memory
access
timeout
The following parameters can be read via a serial interface.
PID Name
Description
Factory
setting
Min.
Max.
Type
221 Pressure
Output pressure in integer format
Fixs32en20
222 Pressure
Pressure in vacuum chamber in Real
format
Real32
265 ATM Pressure
Ambient pressure in real format
Real32
466 Differential Pressure
Differential pressure outside and inside
vacuum chamber
Real32
224 Data Unit
Changes the pressure unit of all Real32
pressures.
0 mbar
1 Torr
2 Pascal
3 micron
4 Counts
0
0
4
Uint8
PID Name
Description
Factory
setting
Min.
Max. Type
228 Device Exception
0 No
error
1
Timeout EEPROM memory access
2
EEPROM CRC error
3 EEPROM
error
4
Pirani filament rupture
5 Wrong
filament
material
6 CDG
diaphragm
rupture
8
ATM outside specification limits
11 Sensor does not match gauge
0
Uint8
Communication Error
Protocol Description
Output pressure
Error