1.2.1 Removing flex

cable

Be very careful when removing the

flexible cable in order not to demage

the rivet, the cable or the circuit

board. Read the whole section 1.2.1

before commencing work.

- First remove the rivet by carefully

pulling it upwards.

- Thereafter pull the holder the same

way.

Figure 1, Rivet and holder in place

Figure 2, Rivet is removed holder

still in place

Figure 3, Rivet and holder removed

Next step before pulling out the flexi-

ble cable is opening the circuit board

connector (see next page).

Rivet

Holder

Connector

Flexible cable

Circuit board

Figure 1